| Model Number |
8575KA |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device in question was returned to the manufacturer.The evaluation is anticipated, but not yet begun.The investigation will be performed by a designated karl storz employee.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the device had been wearing out with some metal powder on it, and the defect was noted after reprocessing procedure and not involving medical procedure" there is no information that the event caused a harm or serious injury to patient, user or third.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: result of investigation: accordingly, we would like to give a summary of our results regarding the reported issue.The investigation of the returned product revealed chip formation at the thread, which confirmed the customer complaint.The event is filed under internal karl storz complaint id: (b)(4).
|
| |
|