| Catalog Number |
MPXT5302-C |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
07/09/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd mpxt5302-c - 7.5 in minibore pressure rated w/t-conn leaked at a connection.The following information was provided by the initial reporter: according to the customer, there have been subjective concerns with the product in up to 70% of power injection attempts.They provided 12 documented examples from staff, all involving ct procedures where leakage occurred during power injection.The leakage was observed either at the connection between the catheter hub and the j-loop extension set or between the j-loop and the maxplus.In this specific case, leakage was reported between the insyte ag catheter hub and the j-loop extension set during a power injection in ct.While no specific injection rates, gage sizes or psi values were provided, the customer confirmed that they do not exceed 325 psi during these procedures.(b)(6) 2025: she has mentioned she doesn¿t have any additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
No product or photo was returned by the customer.The customer complaint of leaking at the j-loop during power injection could not be verified due to the product not being returned for failure investigation.The mpxt5302-c set is rated for power injection according to the ifu up to a pressure of 325 psi.If the needs of the customer are outside of these parameters, please reach out to your bd sales rep for guidance.The root cause for the connection and leaking issue could not be found without a sample investigation.We understand the challenging connection between the spin collar and the male luer component to the catheter hub, and bd is available to help if the customer has any further issues.A device history record review could not be performed because the lot number is unknown.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
|
| |
|
Event or Problem Description
|
|
No additional information was provided.
|
| |
|
Search Alerts/Recalls
|