|
W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
PXC201000 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Aneurysm (1708)
|
| Date of Event |
07/23/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2010, the patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis devices.At an unknown date, the left common iliac artery, at the site where the distal end of a contralateral leg endoprosthesis had been implanted, was found to be dilated.On (b)(6) 2025, a reintervention was performed.The left internal iliac artery was embolized, and a gore® excluder® aaa endoprosthesis (contralateral leg endoprosthesis) was implanted to extend the initially implanted contralateral leg endoprosthesis to the left external iliac artery.The procedure was well tolerated by the patient.The physician stated that the initially implanted contralateral leg endoprosthesis was undersized because the diameter of the largest commercially available device at that time was 20 mm.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|
|
|