| Model Number |
651001DAR |
| Medical Device Problem Code |
Increase in Pressure (1491)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During the mattress inspection, the arjo technician found the faulty automatt.The mattress was repaired by the automatt replacement with a new one.The mattress has been put back in use.The cause of the automatt over-inflation is incorrect circulation of the air in the automatt sensor pad (mattress base) due to inner tube damage.The tpu tube (p/n nmb513) may break over time due to the extruding force of the silicone tube and the external bending force.The membranes of the grommets were punctured.When the grommet membrane is punctured and load is applied on the mattress, air will escape through the punctured membrane, reducing the risk of the automatt over-inflating and the patient's falling.In the complaint at hand, the load was not applied to the mattress (there was no patient involved).This is in line with the instruction for use for nimbus 4 and nimbus professional (649933en): "do not place the patient on the mattress until it is fully inflated and normal operating pressure has been reached." to sum up, the nimbus professional mattress did not meet its specifications since the automatt sensor pad was faulty.No patient was involved when the reported issue occurred.No injury was sustained.The complaint was decided to be reportable due to the nature of the failure (automatt overinflated).The device is distributed outside the us and no gudid information exists.
|
| |
|
Event or Problem Description
|
|
Arjo was informed about an event involving the nimbus professional mattress.The customer reported that the overinflation issue was identified by facility staff while preparing the bed for patient transfer onto it.The issue was detected before any patient interaction.No patient was involved, and no injury was sustained.
|
| |
|
Search Alerts/Recalls
|