• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARJO (SUZHOU) CO., LTD. NIMBUS PROFESSIONAL; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARJO (SUZHOU) CO., LTD. NIMBUS PROFESSIONAL; MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE Back to Search Results
Model Number 651001DAR
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
During the mattress inspection, the arjo technician found that the automatt was faulty (it had inflated like a balloon), and overinflation occurred.The cause of the automatt over-inflation is incorrect circulation of the air in the automatt sensor pad (mattress base) due to inner tube damage.The tpu tube (p/n nmb513) may break over time due to the extruding force of the silicone tube and the external bending force.The membranes of the grommets were punctured.When the grommet membrane is punctured and load is applied on the mattress, air will escape through the punctured membrane, reducing the risk of the automatt over-inflating and the patient's falling.In the complaint at hand, the load was not applied to the mattress (there was no patient involved).This is in line with the instruction for use for nimbus 4 and nimbus professional (649933en): "do not place the patient on the mattress until it is fully inflated and normal operating pressure has been reached." to sum up, the nimbus professional mattress did not meet its specifications since the automatt sensor pad was faulty.No patient was involved when the reported issue occurred.No injury was sustained.The complaint was decided to be reportable due to the nature of the failure (automatt overinflated).The device is distributed outside the us and no gudid information exists.
 
Event or Problem Description
Arjo was informed about the event related to the nimbus professional mattress.The customer reported a malfunction of the mattress, indicating that it no longer inflates as expected.There was no indication of patient involvement.No injury was sustained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIMBUS PROFESSIONAL
Common Device Name
MATTRESS, AIR FLOTATION, ALTERNATING PRESSURE
Manufacturer (Section D)
ARJO (SUZHOU) CO., LTD.
no. 158 fangzhou road, sip
suzhou, jiangsu 21502 4
CH  215024
Manufacturer (Section G)
ARJO (SUZHOU) CO., LTD.
no. 158 fangzhou road, sip
suzhou, jiangsu 21502 4
CH   215024
Manufacturer Contact
justyna kielbowska
ks. wawrzyniaka 2
komorniki 62-05-2
PL   62-052
883337089
MDR Report Key22713581
Report Number3005619970-2025-00031
Device Sequence Number9628041
Product Code FNM
UDI-Device Identifier05055982782197
UDI-Public(01)05055982782197(11)160421
Combination Product (Y/N)N
Initial Reporter CountryBE
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number651001DAR
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2025
Initial Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/21/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
RES# 87752
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-