• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. VIDEO SYSTEM; VIDEO SYSTEM CENTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. VIDEO SYSTEM; VIDEO SYSTEM CENTER Back to Search Results
Model Number OTV-SI
Medical Device Problem Codes No Display/Image (1183); Loose or Intermittent Connection (1371)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported that the output connector port is loose and that there is no image on video system device.The issue occurred during setup.There is no report of any patient harm.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d9, h2, h3, h4, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: worn-out connector was causing the connection problem.The most probable cause of the complaint is traced to component failure.Since the device is more than 15 years old, the device history record was unable to be reviewed.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information received from customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIDEO SYSTEM
Common Device Name
VIDEO SYSTEM CENTER
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22713583
Report Number3002808148-2025-12155
Device Sequence Number9628042
Product Code FET
UDI-Device Identifier04953170169885
UDI-Public04953170169885
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K954451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 09/03/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOTV-SI
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/08/2025
Supplement Date Received by Manufacturer08/06/2025
Initial Report FDA Received Date08/05/2025
Supplement Report FDA Received Date09/03/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/25/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-