Catalog Number 21-7001-24
Medical Device Problem Codes Excess Flow or Over-Infusion (1311); Inaccurate Delivery (2339)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/10/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H3: investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event or Problem Description
It was reported that the total volume was 153 ml.The pump finished 8 hours early saying there was still 53 ml in cassette.The cassette was empty.The amount of medication remaining in cassette/bag did not match the reservoir volume displayed on pump.There was patient involvement, but no patient harm was reported.
 
Additional Manufacturer Narrative
H3: the device was received for evaluation.The lot history was reviewed; no nonconformities were identified that may have contributed to the reported complaint.Visual and functional tests were performed.The customer's stated problem was not confirmed.The device was delivering properly.There was no known cause of the reported issue, and no further action will be taken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD MEDICATION CASSETTE
Common Device Name
PUMP, INFUSION
Manufacturer (Section D)
ICU MEDICAL, INC.
ave calidad no. 4, parque
tijuana
MX 
Manufacturer Contact
reed covert
6000 nathan lane north
minneapolis, MN 55442
2247062300
MDR Report Key22714218
Report Number3012307300-2025-09294
Device Sequence Number15801165
Product Code FRN
UDI-Device Identifier10610586022111
UDI-Public(01)10610586022111(17)290611(10)6022008
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K843772
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/29/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number21-7001-24
Device Lot Number6022008
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/28/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2025
Supplement Date Received by Manufacturer09/11/2025
Initial Report FDA Received Date08/05/2025
Supplement Report FDA Received Date09/29/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1