| Model Number |
CCWTT0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Halo (2227)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-healthcare professional reported that following an intraocular lens (iol) implant procedure, the patient experienced halos.The iol was exchanged for unspecified lens model 2 months and 22 days after the initial implant procedure.Clinical reason for explant was reported as intolerance of halos.Additional information has been requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in b.2., b.5.And h.11.Based on information received following submission of the initial report, this event does not meet criteria for reporting as a serious injury.No further reports required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional information received and reported that the iol was exchanged for the same company lens with different model but half a diopter less power indicating the correct lens power was not implanted initially.There is no reported defect or fault with the iol.
|
| |
|