| Catalog Number |
122136058 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
07/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: the expiration date is currently not available.Therefore, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient is three weeks post op and developed an infection.Patient was brought back to the or for an i&d with head and poly exchange.Doi: (b)(6) 2025.Dor: (b)(6) 2025.Affected side: left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A search of the medtech orthopedics nonconformance (nc) quality system found no nc¿s associated with this product<122136058>/lot<4473225> combination.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: a search of the medtech orthopedics nonconformance (nc) quality system found no nc¿s associated with this product<122136058>/lot<4473225> combination.Added: d4 (expiry date), d10 concomitant, h4.Corrected: d4 primary udi number.
|
| |
|
Search Alerts/Recalls
|