| Catalog Number |
242107 |
| Medical Device Problem Code |
Display or Visual Feedback Problem (1184)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
|
| |
|
Event or Problem Description
|
|
It was reported that during an unspecified surgical procedure it was observed that the hd epscp,2.7,30,75,mitek scope device was displaying a cloudy image.It was reported that the device was swapped for another device and the procedure was successfully completed.There were no adverse consequences to the patient.No additional information could be provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer.Outer tube damaged, needle outer tube dent(s) deep dents outer tube bent.Outer tube damaged, distal tip distal tip damaged.Illumination distal tip fiber damaged.Optical system, optical components broken lenses in optical system.Cosmetic issue scratches/dents.Labeling : illegible etch, visual : deformed/bent, broken (2+ pieces) : chipped/shaved/delaminated, visual : scratched/nicked.Per service reports, this complaint was confirmed.The defective parts need to be replaced to resolve the issues.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review a manufacturing record evaluation was performed for the finished device serial number, and no non-conformance was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: during complaint investigation it was determined that the device evaluation required an update.Investigation summary: the complaint device was received at the service center and evaluated.During the service evaluation the following defects were identified: other damages caused by customer.Outer tube damaged, needle outer tube dent(s) deep dents outer tube bent.Outer tube damaged, distal tip distal tip damaged.Illumination distal tip fiber damaged.Optical system, optical components broken lenses in optical system.Cosmetic issue scratches/dents.Labeling: illegible etch, visual: deformed/bent, broken (2+ pieces): chipped/shaved/delaminated, visual: scratched/nicked.Engraving/identification datamatrix code level a.Per service report, this complaint was confirmed.The faulty parts were replaced to resolve the issues.After repair, the device was found to be working according to the specifications.Based on the investigation findings, the potential root cause is traced to user error.It has been determined that no corrective and/or preventative action is proposed and there is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: a manufacturing record evaluation was performed for the finished device serial number, and no non-conformance was identified.
|
| |
|