ARTHREX, INC. ANGLED REAMER, DRIVE SHAFT; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number ANGLED REAMER, DRIVE SHAFT
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 07/16/2025, it was reported by a sales representative via (b)(4) that an ar-9597-20 angled reamer sleeve was cold welded with inside of the reamer.This was discovered during the case with no patient harm.Additional information was received on 7/21/25, that this was discovered during a reverse total shoulder arthroplasty.After the 20 degree mgs augment reamer malfunctioned, the 25 degree primary augment reamer was used to complete reaming.Additional information was provided, on 7/24/25, that the sleeve was stuck to the drive shaft.Part number and sn provided.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the returned device (if available), pictures, videos, event description, and any additional field data, arthrex concluded that the most likely cause of the reported failure.The most likely cause for the reported failure can be attributed to misuse due to a force perpendicular to the drive shaft being applied while the device is spinning.This applied force creates an excess of friction and heat that ultimately has the potential to cause the device to cease functioning as intended.Application of this force perpendicular to the drive shaft is not needed for the device to function, nor is it recommended that the user does this during normal clinical use.When used usually, this issue is unlikely to present itself, which suggests that the handling of the device greatly impacts whether or not it will seize.For this reason, the direct cause is considered to be device misuse.The complaint allegation could not be confirmed, as the device was not received for evaluation, and no evidence of the reported failure was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANGLED REAMER, DRIVE SHAFT
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22717568
Report Number1220246-2025-03251
Device Sequence Number18443235
Product Code LXH
UDI-Device Identifier00888867324954
UDI-Public00888867324954
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/10/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberANGLED REAMER, DRIVE SHAFT
Device Catalogue NumberAR-9676
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/16/2025
Supplement Date Received by Manufacturer11/04/2025
Initial Report FDA Received Date08/05/2025
Supplement Report FDA Received Date11/10/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-