• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US ULTRA AGGRESSIVE PLUS 4.0MM 5PK; ARTHROSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US ULTRA AGGRESSIVE PLUS 4.0MM 5PK; ARTHROSCOPE Back to Search Results
Catalog Number 283429
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to an arthroscopy surgical procedure it was observed that the ultra aggressive plus 4.0mm 5pk device leaked.There was a milky white lubricant inside the planer blade which solidified and flowed out.Another like device was used to complete the procedure.There were no reports of injuries, medical intervention or prolonged hospitalization.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the product was not returned to j&j medtech orthopaedics, however a photo was provided for evaluation.Visual inspection of the returned photo found that the device is shown in used condition.The tip of the device shows the white lubricant that is used for the proper rotating function.A previous investigation with the manufacturer brought that there are controls in place related to the grease application.The outer tip grease, and in the line release, the quality inspector takes one piece during the shift or lot production and weight the grease of the inner/outer based on the quality work instruction, the results are documented.Also, there is 100% visual inspection during the process to review the final condition of the lubricant in the blade.On final inspection and cleaning process, a step of the inspection is to verify that there is no grease excess on the tip; this is an important control that avoid quality rejection.The overall complaint was not confirmed as the observed condition of the ultra aggressive plus 4.0mm 5pk would not contribute to the complained device issue.Based on the investigation findings, the potential cause is traced to procedural variables, such handling of the device or product interaction during procedure; during the installation of the blade, the operator may pulled the inner shaft out and inserted again, causing the lubricant to allocate at the tip.However, this cannot be conclusively determined.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of j&j medtech orthopaedics quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRA AGGRESSIVE PLUS 4.0MM 5PK
Common Device Name
ARTHROSCOPE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key22717595
Report Number1221934-2025-02974
Device Sequence Number13053792
Product Code HRX
UDI-Device Identifier10886705022229
UDI-Public(01)10886705022229(10)M2411007(17)291031
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K131191
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number283429
Device Lot NumberM2411007
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/23/2025
Initial Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/18/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-