• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENTRANT¿ DR; Implantable cardioverter defibrillator (non-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ENTRANT¿ DR; Implantable cardioverter defibrillator (non-CRT) Back to Search Results
Model Number CDDRA300T
Medical Device Problem Codes Over-Sensing (1438); Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/19/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented in-clinic as the implantable cardioverter defibrillator (icd) and right-ventricular (rv) lead exhibited inappropriate shock as a result of oversensing electromagnetic interference of another implanted device.The rv lead exhibited low pacing impedance as well.Programming changes were made initially to deactivate the icd.Additionally, upon fluoroscopy imaging it was noted that the rv lead exhibited externalized conductors, and noise was reproducible by manipulating output frequencies of the other device.As a resolution the rv lead and the icd were explanted and replaced.The patient was recovering.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTRANT¿ DR
Common Device Name
Implantable cardioverter defibrillator (non-CRT)
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key22719156
Report Number2017865-2025-96779
Device Sequence Number9393346
Product Code LWS
UDI-Device Identifier05415067049469
UDI-Public(01)05415067049469(10)S000097777(17)260731
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/05/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberCDDRA300T
Device Catalogue NumberCDDRA300T
Device Lot NumberS000097777
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/19/2025
Initial Report FDA Received Date08/05/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/30/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BAROSTIM.; ENTRANT.; SDX.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age62 YR
Patient SexMale
Patient Weight77 KG
-
-