MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SSF-770
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 07/09/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture baker grade iv" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
 
Event or Problem Description
Healthcare professional reported "capsular contracture baker grade iv".This record is for the right side.Device remains implanted.
 
Additional Manufacturer Narrative
Reason for reoperation: capsular contracture baker grade iv, rupture.
 
Event or Problem Description
Healthcare professional later reported "hole on patch side".Device has been explanted.
 
Event or Problem Description
Healthcare professional reported "capsular contracture baker grade iv".Later healthcare professional reported "right side is tight deformed and pushed up and painful.Therefore [patient] has a baker grade 4 capsular contracture".Later healthcare professional reported "hole on patch side".This record is for the right side.The device was explanted.
 
Additional Manufacturer Narrative
Lab analysis results: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.¿ rupture: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis (non-penetrating nicks, creases) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
8479366324
MDR Report Key22720006
Report Number9617229-2025-13209
Device Sequence Number9393695
Product Code FTR
UDI-Device Identifier10888628033252
UDI-Public(01)10888628033252(10)3354739(11)190627(17)240527
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/27/2024
Device Catalogue NumberSSF-770
Device Lot Number3354739
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/09/2025
Supplement Date Received by Manufacturer03/13/2026
04/27/2026
Initial Report FDA Received Date08/05/2025
Supplement Report FDA Received Date04/03/2026
05/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/27/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-