ORBUSNEICH MEDICAL (SHENZHEN) CO., LTD JADE 18 OTW PTA BALLOON DILATATION CATHETER; Catheter, angioplasty, peripheral, transluminal Back to Search Results
Catalog Number 596015022
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/11/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during procedure, a jade balloon was placed in patient's superficial femoral artery (sfa) and was inflated to 12 atmospheres.The balloon ruptured.During removal, it was noticed that the balloon had broken in-vivo and became completely detached.An attempt was made to advance another wire through the existing 6f non-abbott contralateral sheath, but was unsuccessful.The physician had to puncture the sfa antegrade directly below the broken balloon.A stent was placed with difficulty to adhere the balloon against the wall.There was delay in the treatment as it added 75 to 90 minutes to the procedure with two additional punctures that were required.The patient was discharged after bedrest.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JADE 18 OTW PTA BALLOON DILATATION CATHETER
Common Device Name
Catheter, angioplasty, peripheral, transluminal
Manufacturer (Section D)
ORBUSNEICH MEDICAL (SHENZHEN) CO., LTD
no.1st jinkui road
futian free trade zone
shenzhen, guangdong 51803 8
CH  518038
MDR Report Key22724080
Report Number3014148220-2025-00272
Device Sequence Number8344514
Product Code LIT
UDI-Device Identifier06934955903939
UDI-Public(01)06934955903939(11)230728(17)250728(10)5930132307
Combination Product (Y/N)N
Initial Reporter StateNE
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number596015022
Device Lot Number5930132307
Was Device Available for Evaluation? Yes
Distributor Facility Aware Date07/11/2025
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/06/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
-
-