STRYKER ORTHOPAEDICS-MAHWAH TI DUR REG FLUTED STEM13X155MM; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
| Catalog Number |
64786695 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
05/11/2013
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event regarding infection involving a duracon stem was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional, and dimensional inspections could not be performed as the device remains implanted -clinician review: no medical records were received for review with a clinical consultant.-product history review: could not be performed as the device lot details were not provided.-complaint history review: could not be performed as the device lot details were not provided.Conclusions: all stryker products sold as sterile are validated to a minimum sterility assurance level (sal) of 10^-6 in accordance to applicable iso standards.The exact cause of the event could not be determined because insufficient information was provided.Additional information including pathology reports, operative reports, progress notes, x-rays and return of the device are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.The following devices were also listed in this report: cat# 64952010; gmrs dist fem comp sml l 65mm; lot# unknown.Cat# 64813110; mrh tibial b/plt keel sml; lot# unknown.It cannot be determined which, if any of these devices may have caused or contributed to the patient's experience.
|
| |
|
Event or Problem Description
|
|
This pi is for the wound infection.Information received from (b)(4) indicates the following: wound infection suspected (b)(6) 2013, iv vanc, no return to or.Later needed to be on lifelong oral antibiotics for chronic suppresion.
|
| |
|
Search Alerts/Recalls
|
|
|