• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; SILICONE FOLEY TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD INC. (COVINGTON) -1018233 LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY; SILICONE FOLEY TRAY Back to Search Results
Catalog Number 319214A
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during the pre-test, it became difficult to drain the water from foley catheter, and it became unusable.
 
Additional Manufacturer Narrative
The reported event is confirmed via capa associated.This is addressed by capa 12474540.Received 1 all silicone foley catheter with syringe.Verified material number 119314 and manufacturing lot number ngjx0304.Visual inspection noted no obvious observations.During testing, the catheter filled with 10ml methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) using returned syringe, the balloon inflated with no difficulty.Concentricity of catheter balloon is 89/11.Only 9 ml of methylene blue solution in the catheter balloon was deflated within 5 minutes.Again, the catheter was filled with 10 ml of methylene blue solution (3 drops 1% aq methylene blue per 100ml distilled water) using an in house syringe.The balloon deflated 10ml of solution within 02min09sec.This is out of specification per inspection procedure (b)(4) which states, catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.The reported event is confirmed against the returned syringe.Capa 12474540 identified the root cause of this failure as a combination of three (3) factors; provided water syringe does not meet user expectation for smooth engagement with the valve and no instructions are provided in the japanese ifu for aspirating to assist in deflation, deflation time or take off force criteria or specification to deflate the catheter has not been determined and it is unknown for the customers, inadequate process qualification performed with change in syringe supplier and mold change.Risk review, labeling review, and dhr review are not required as event has already been investigated per capa 12474540.Refer to capa 12474540 for further details.Correction: d, e, f, h.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during the pretest, it became difficult to drain the water from foley catheter, and it became unusable.Per notification from investigator on 12jan2026, it was reported that asymmetrical balloon was found during sample evaluation on 15dec2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUBRI-SIL® I.C. COMPLETE CARE® TEMPERATURE-SENSING URINE METER FOLEY TRAY
Common Device Name
SILICONE FOLEY TRAY
Manufacturer (Section D)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key22727583
Report Number1018233-2025-06521
Device Sequence Number15525800
Product Code MJC
UDI-Device Identifier00801741040153
UDI-Public(01)00801741040153(11)241126(17)291120(10)NGJX0304
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K070582
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number319214A
Device Lot NumberNGJX0304
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2025
Supplement Date Received by Manufacturer03/19/2026
Initial Report FDA Received Date08/06/2025
Supplement Report FDA Received Date03/21/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/26/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-