The device referenced in this report was not returned to shockwave medical, inc.Therefore, a physical examination cannot be performed.The cause of the perforation and cardiac arrest could not be definitively determined based on the information available.Balloon ruptures are a known failure mode with a low probability of occurrence.Contributing factors include patient calcium, damage caused when the balloon wall comes into close contact with the device emitters, likely causes include (a) an irregular calcium lesion capable of compressing the balloon wall into close proximity with the emitter(s) and/or (b) an incompletely inflated balloon.The shockwave intravascular lithotripsy (ivl) system with the shockwave c2+ coronary ivl catheter generates acoustic pressure pulses within the target treatment site, disrupting calcium within the lesion allowing subsequent dilatation of a coronary artery stenosis using low balloon pressure not to exceed 4 atm during ivl treatment, and 6 atm for post dilatation.The pressure used for the ivl balloon catheter is much lower than other balloon catheters used in percutaneous coronary intervention (pci), thus the risk associated with loss of balloon pressure during ivl catheter use is negligible.The associated patient risk for this failure mode is low and adequately captured in swmi's risk documentation.Shockwave medical has controls in place to ensure devices are built to approved procedures and meet lot release acceptance criteria prior to being distributed.A review of the manufacturing and test documentation for subject lot does not reveal any issues with the manufacturing of the device.The device passed all of shockwave medical, inc.Acceptance criteria prior to shipping.
|
|
A shockwave c2+ coronary intravascular lithotripsy (ivl) catheter was used to treat a lesion in the coronary artery.During the procedure, 30 pulses were delivered prior to the ivl balloon losing pressure.The physician subsequently removed the ivl device and proceeded with a non-compliant (nc) balloon.The patient experienced cardiac arrest and was resuscitated with cpr.Following the ivl treatment, the artery was perforated by the stent.A covered stent was successfully implanted to address the perforation.The patient was stabilized and transferred to the icu.
|