• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR INC. NC TREK NEO¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR INC. NC TREK NEO¿; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 1400550-12
Medical Device Problem Codes Material Rupture (1546); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.Production record and corrective and preventative actions (capa) reviews were performed and revealed no indication of a product quality issue.Additionally, a review of the complaint history identified no other similar complaints from this lot.Based on the information received, the investigation determined that the reported complaints appear to be related to operational context.There was no leak noted to the balloon during preparation, which suggests a product quality issue did not contribute to the reported difficulties.It was reported by the account that the balloon dilatation catheter (bdc) interacted with the moderately tortuous and heavily calcified anatomy resulting in the reported difficulty advancing the device and subsequently the balloon became damaged and ruptured during inflation.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a heavily calcified and moderately tortuous lesion in the left anterior descending artery (lad).After a 5.5x12mm nc trek neo balloon dilatation catheter (bdc) was prepared with no noted issues, the bdc was advanced to the lesion, through resistance, and attempted to be used for post-dilatation.However, during the initial inflation, the balloon was noted to rupture at 10atms, therefore the bdc was replaced with another balloon catheter and the procedure was completed.There were no adverse patient effects nor clinical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK NEO¿
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
ABBOTT VASCULAR INC.
3200 lakeside drive
santa clara CA 95054 2807
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela UNK
CS   UNK
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key22729066
Report Number2024168-2025-08117
Device Sequence Number18214844
Product Code LOX
UDI-Device Identifier08717648232831
UDI-Public(01)08717648232831(17)270831(10)40911G1
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K220634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/06/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1400550-12
Device Catalogue Number1400550-12
Device Lot Number40911G1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2025
Initial Report FDA Received Date08/06/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/11/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-