| Catalog Number |
22-4038 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an arthroscopic procedure, the firstpass functioned for several passes without complication, but later had resistance when passing the device.The surgeon attempted to proceed but was unable to complete the pass.The device was then carefully withdrawn from the joint, and it was discovered that the central core of the firstpass device had broken into two pieces.One half of the broken piece was missing.As precaution, an intraoperative x-ray was performed to assess for any retained foreign objects within the joint, but the imaging revealed no visible remnants left inside the patient.Surgery was completed with a back-up device instead.There was a surgical delay of less than 30 minutes, and no further complications were reported.Patient is recovering well.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known inherent device and/or procedural risks that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are conservatively submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: the reported device was received for evaluation.A visual evaluation showed the device was not returned with any original packaging.The suture capture has been disconnected from the upper jaw.A partial suture capture was returned.Bio debris is present.No other visual deficiencies.A functional evaluation showed pulling the lever will close the aligned jaws and deploy the suture passer needle as intended.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences ((1) excessive force, (2) tissue thickness, (3) damage or debris on the device tip between passes).Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|
|