| Model Number |
2110 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: the device was received for evaluation.Service history review identified there was no indication that the complaint was related to a service of the device within the review period.Visual and functional tests were performed, which found a lifting downstream occlusion (dso) seal.The dso seal was replaced to fix the issue.
|
| |
|
Event or Problem Description
|
|
The device was returned due to the power port not recognizing the ac adapter.The results of the investigation found that the device's downstream occlusion (dso) seal was lifting.There was no patient involvement.
|
| |
|