| Model Number |
G146 |
| Medical Device Problem Codes |
Premature Discharge of Battery (1057); High impedance (1291); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/10/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device had 2 years of battery life remaining.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Technical services (ts) consultant recommended replacement.This crt-d was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device had 2 years of battery life remaining.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Technical services (ts) consultant recommended replacement.This crt-d was explanted and replaced.No additional adverse patient effects were reported.Additional information was received that this cardiac resynchronization therapy defibrillator (crt-d) had recorded high out of range pacing impedances on the left ventricular (lv) channel prior to explanation.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device had 2 years of battery life remaining.A request was made to have data from this device analyzed.Data analysis confirmed the battery appeared to be depleting more quickly than expected.Technical services (ts) consultant recommended replacement.This crt-d was explanted and replaced.No additional adverse patient effects were reported.Additional information was received that this cardiac resynchronization therapy defibrillator (crt-d) had recorded high out of range pacing impedances on the left ventricular (lv) channel prior to explanation.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.External visual inspection of the device revealed no anomalies.Review of the device memory confirmed that a low voltage alert was recorded.The battery voltage level upon receipt in the laboratory was sufficient to ensure therapy availability/delivery while the device was implanted.The device case was then opened to facilitate analysis of the internal components.The battery was separated from the other device electronics and the overall current draw of the circuitry was measured.A normal current drain was observed within the circuitry.Detailed battery analysis determined that a latent current leakage path had occurred within the battery itself, resulting in a partial depletion of the battery.This behavior is caused by a latent electrical leakage path that develops over time between the anode and cathode within the device battery.In certain circumstances, lithium metal ions can re-plate to form clusters that may result in an internal current leakage.
|
| |
|