| Model Number |
357495 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/11/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The returned device was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation revealed a fine jet of water emerging from the distal portion of the balloon when attempting inflation.Microscopic analysis of the balloon surface showed a small pinhole about 10 mm proximal to the distal x-ray marker.At the origin of the pinhole a deep scratch was observed which has likely been caused by a hard, sharp-edged object such as e.G., anatomical structure.Review of the production documentation confirmed that the device was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each device is tested for air tightness by means of a pressure test and a helium leak test.We can therefore confirm that the device was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patient's anatomy (i.E., calcified target lesion).
|
| |
|
Event or Problem Description
|
|
Passeo-28 balloon catheters were selected for treatment of a moderately calcified lesion (70% stenosis degree) in a moderately tortuous part of the mid sfa.Two balloons ruptured.They could not even be inflated up to maximum pressure before the balloons ruptured at 8 atm.This refers to 1st balloon, the other balloon is reported separately.
|
| |
|
Search Alerts/Recalls
|