• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP; DORO® QR3 SKULL CLAMP TEFLON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP; DORO® QR3 SKULL CLAMP TEFLON Back to Search Results
Model Number 3003-009
Medical Device Problem Code Device Slipped (1584)
Health Effect - Clinical Code Laceration(s) (1946)
Date of Event 06/30/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Customer informed us on (b)(6) that one of our products was involved in a case in which the patient sustained a laceration.
 
Additional Manufacturer Narrative
It is not assumed that the detected deviation (openlock mechanism cannot be opened) contributed to the described event, as it would not have been possible to use the product in this condition.The customer was asked several times to send in the missing components of the skull clamp (torque screw, extension arm) for examination in order to rule out or verify any causal relationships.The parts have not yet been sent in.If the remaining components arrive or any new further findings emerge during this investigation, the relevant information will be submitted in a follow-up report.Our experience is, that the pinning technique is a major factor in terms of the occurrence probability of slippages.We refer to our corresponding recommendations described the product's instruction manual (secure the patient's head to the skull clamp): "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP
Common Device Name
DORO® QR3 SKULL CLAMP TEFLON
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 86
freiburg, baden-wuerttemberg 79111
GM  79111
MDR Report Key22735892
Report Number3003923579-2025-00018
Device Sequence Number9240645
Product Code HBL
UDI-Device Identifier04250435500781
UDI-Public04250435500781
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/09/2025
Report Date (Section F) 07/09/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number3003-009
Device Catalogue Number3003-009
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2025
Distributor Facility Aware Date07/09/2025
Device Age4 MO
Event Location Hospital
Date Report to Manufacturer07/09/2025
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/07/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-