MAQUET SAS VOLISTA; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARDVCS209015A
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
We were informed of the incident by the hospital's clinical engineer.Phone number: (b)(6).Getinge became aware of an issue with one of surgical lights - volista access.Customer reported that a screw in the spotlight's structure fell out during a surgical procedure, and that the screw fell onto the patient.The patient was not harmed by the incident.The screw was retightened to ensure its fixation.The screw is placed in the spring arm and holds the safety ring and the spotlight dome.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.Based on the information collected, it was established that when the event occurred, the surgical lights did not meet its specification, since detachment of the screw from spring arm cover could be considered as technical deficiencies, and in this way the device contributed to event.A root cause analysis was performed by subject matter experts at the manufacturing site.As they stated it was detected that the screw on the spring arm cover detached.This screw is responsible for attaching the cover to the spring arm structure.The manufacturer of the spring arms, ¿ondal¿, led several tests on a spring arm performing 20.000 cycles of up and down movement.The test with a loosened screw, with a complete turn back, shows that it is the only case where the screw continue to loosen.Based on the analysis of the cases reported, several cases were detected, all delivered in the same hospital, on a period less than 1 year after the installation, the other cases were detected more than 1 year after the installation.The cases detected after more than 1 year after the installation correspond to a lack of inspection during the yearly preventive maintenance.To prevent the loosening and the fall of the screw, the technical manuals indicate to check all the fixing screws and to check the hold of the spring arm covers during yearly preventive maintenance or daily inspections.The screw has a long threading, its loosening is visually detectable during cleaning or during yearly preventive maintenance.As improvement measure, since october 2019, the spring arms covers are assembled, on the product line, with tuflok coated screws.In any case, by design, the nylon patch of this screw acts as a mechanical locking and prevents the loosening and the fall of the screw.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - volista access.Customer reported that a screw in the spotlight's structure fell out during a surgical procedure, and that the screw fell onto the patient.The patient was not harmed by the incident.The screw was retightened to ensure its fixation.The screw is placed in the spring arm and holds the safety ring and the spotlight dome.We decided to report the issue in abundance of caution as any parts falling off into sterile field or during procedure may cause contamination or serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VOLISTA
Common Device Name
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key22736481
Report Number9710055-2025-0000200
Device Sequence Number12562381
Product Code FSY
UDI-Device Identifier03700712420475
UDI-Public(01)03700712420475
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K130513
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberARDVCS209015A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/24/2025
Initial Report FDA Received Date08/07/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/21/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-