| Catalog Number |
103545000 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as a valid lot number for the device involved in the event was not provided, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
It was reported that the liner would not fit the bipolar shell, it was too loose and would not seat in the component.Affected side: unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H11 additional narrative: added: a2 (age), d4 (lot, expiry date), h4.Corrected: d4 (primary udi number).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4) investigation summary: no device associated with this report was received for examination.According to the information received, ¿it was reported that the liner would not fit the bipolar shell, it was too loose and would not seat in the component.¿ product description: self cent hip 45x28 gry.Product code: 103545000.Lot no: d25031784.Quantity of manufactured: (b)(4).Date of manufacturing: 06-mar-2025.Expiry date: 28-feb-2030.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: product description: self cent hip 45x28 gry.Product code: 103545000.Lot no: d25031784.Quantity of manufactured: (b)(4).Date of manufacturing: 06-mar-2025.Expiry date: 28-feb-2030.Device history review: a manufacturing record evaluation was performed for the finished device (b)(4), and no non-conformances / manufacturing irregularities were identified.H11 additional narrative: added: d10.
|
| |
|
Search Alerts/Recalls
|