This supplemental report is being submitted to provide the results of the final investigation.Updated fields: b5, d8, d9, h2, h3, h6, h11.The device was returned to olympus for inspection and the reported failure was confirmed.In addition, the following reportable malfunctions were identified during the device evaluation: the light guide bundle was chipped, caused by a component failure of the light guide bundle.A root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: caused by a component failure of the universal cable.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|