A healthcare professional reported that prior to use, the doctor opened the device packaging of an enterprise2 4mmx23mm vascular reconstruction device (encr402312, 9061888) and removed the device from dispenser hoop.It was found that the stent was released automatically.The stent was detached from the delivery wire.The device was not used in the patient.The doctor switched to a new stent to complete the surgery.Additional event information received on 24-jul-2025 indicated that there was no other physical damage to the device besides the reported premature detachment.The replacement stent was another enterprise2 4mmx23mm® vascular reconstruction device.The event did not result in a clinically significant delay in the procedure.
|
|
Manufacturer ref#: (b)(4).Since no device has been received for analysis, no product investigation can be performed and the reported failure could not be evaluated.A device history record (dhr) was performed and it indicated this product was final inspection tested at lake region medical and was determined to be acceptable.With the information available and without the product available for analysis, the reported customer complaint could not be confirmed.Based on the device history record review, there is no indication that the event is related to the device manufacturing process.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|