• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 PIPELINE FLEX W/SHIELD TECHNOLOGY; INTRACRANIAL ANEURYSM FLOW DIVERTER Back to Search Results
Model Number PED2-400-12
Medical Device Problem Codes Retraction Problem (1536); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Continuation of d10: product id fg15150-0615-1s (b)(6); product type: implant date n/a; explant date n/a medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received a report that the patient was undergoing treatment for a saccular, unruptured internal carotid artery (ica) paraophthalmic aneurysm with a max diameter of3.5 mm and a 3.0mm neck diameter.It was noted that the patient's vessel tortuosity was moderate.The landing zone was 3.3mm distally and 3.9mm proximally.The access vessel was groin, with 4-6mm diameter.It was reported that after deploying the pipeline ped, the physician was unable to pull the sleeves into the phenom 027 catheter and encountered resistance at the distal section.The angiographic result post procedure showed a good result; the device was well apposed.The pipeline did not get stuck, and the catheter and pushwire were not damaged.The pipeline was used for an indication that is approved (on-label).The reported device and any accessory devices were prepared and flushed as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.Ancillary devices include a synchro 014 guidewire, phenom 27 microcatheter, sophia ex 6fr guide catheter.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Additional information was received stating that the device was prepared as indicated in the ifu.The procedure was completed by pulling wire and catheter out together without retrieving sleeves.The information is confirmed by the physician.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PIPELINE FLEX W/SHIELD TECHNOLOGY
Common Device Name
INTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer (Section G)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
7635265677
MDR Report Key22738515
Report Number2029214-2025-01773
Device Sequence Number20630323
Product Code OUT
UDI-Device Identifier00763000284329
UDI-Public(01)00763000284329(17)271031(10)B812100
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P100018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPED2-400-12
Device Catalogue NumberPED2-400-12
Device Lot NumberB812100
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2025
Supplement Date Received by Manufacturer08/14/2025
Initial Report FDA Received Date08/07/2025
Supplement Report FDA Received Date09/08/2025
Date Device Manufactured10/31/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient SexFemale
-
-