| Model Number |
ACU0T0 |
| Medical Device Problem Code |
Failure to Fold (1255)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A health care professional reported that during an intraocular lens (iol) implant procedure, the lens became twisted after being advanced and made contact with the patient.The procedure was completed on the same day with an alternative lens implanted successfully.No patient harm occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A non-qualified viscoelastic was indicated.The root cause may be related to a failure to follow the instructions for use (ifu).A non-qualified viscoelastic was used in the device.The ifu instructs: during device preparation and implantation of the company product, a company qualified ophthalmic viscoelastic device (ovd) should be used.The use of an unqualified ovd may cause damage to the lens and potential complications during the device preparation and implantation steps.The account indicated the product was not available to evaluate.The ifu instructs: take at least 7 seconds to gently fold the intraocular lens (iol) by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the "fill-to" line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.After the lens has been advanced to the "fill-to" line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.No part of the lens should exit the nozzle prior to insertion through the incision.Once the lens is in position at the "fill-to" line, it should be implanted within 3 minutes.Proceeding with implantation of a misfolded haptic or a lens that appears to be "out of position" can result in a broken haptic or other negative outcome.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|