Model Number 0171
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/24/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that this right ventricular (rv) lead recorded high out of range shock impedance measurements.Technical services (ts) reviewed the lead data and noted that the lead trend was around 100 ohms.Ts discussed the advisory and provided reprogramming recommendations.No adverse patient effects were reported.This rv lead remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE SG
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key22740237
Report Number2124215-2025-55140
Device Sequence Number9104284
Product Code LWS
UDI-Device Identifier00802526373374
UDI-Public(01)00802526373374(17)181027
Combination Product (Y/N)Y
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P910073
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/07/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/27/2018
Device Model Number0171
Device Catalogue Number0171
Device Lot Number104727
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/25/2025
Initial Report FDA Received Date08/07/2025
Date Device Manufactured10/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
-
-