It was reported that during a thoracoscopic esophagectomy, gcs40b was used for the gastric tube angioplasty.When taking out the product from the package before use, the push rod had already come off.The doctor was concerned about a miss-firing, so the device was refrained from using, and a new gcs40b was taken out and used instead.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
(b)(4).B3: unknown; captured as awareness date.Date sent 8/7/2025.D4 batch #262d81.Investigation summary the product was returned for thorough evaluation, which included visual inspections and functional testing of the returned device.Visual analysis of the returned sample determined that the gcs40b device was received with no apparent damage and with a reload loaded in the device.The reload was received fully loaded with staples, with the washer uncut, with the drivers and with the knife recess below the reload deck.The reload was not properly loaded in the device, as the retaining pin was not connected to the coupler and pushrod.These facts indicate that the reload was removed and reinserted incorrectly and/or incompletely after the retainer was removed.The returned reload was pulled out of the device and placed back on; the device opened and closed without any difficulties.The device was tested for functionality with an returned reload and the device fired, forming all staples and cutting as intended.The cut line and the staple line were completed and the staples meet the staple form release criteria.It should be noted that if the reload is removed from the device, whether the reload is spent or not, and if the staple retainer has already been removed from the reload, the reload cannot be reloaded into the device.Please reference the instructions for use for the complete guide loading and reloading the instrument.As part of ethicon endo surgery¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 262d81, and no non-conformances were identified.
|