| Model Number |
DLPM35MN02 |
| Medical Device Problem Codes |
Device Alarm System (1012); Image Display Error/Artifact (1304); Charging Problem (2892)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, the unit did not charge and had the error message ¿check sum error¿ produced without sounding an alarm.Troubleshooting steps included power cycling the unit, replacing the power cord, and swapping the batteries to isolate the issue.There was no patient involvement.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to use, the unit did not charge and had the error message ¿check sum error¿.Troubleshooting steps included power cycling the unit, replacing the power cord, and swapping the batteries to isolate the issue.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b5, h6 (removed rfr code no alarm) additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned.Therefore, this event does not meet the reporting requirements in 21 cfr 803.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|