| Catalog Number |
LONG60A |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 8/7/2025.D4: batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Was the device locked on tissue with the jaws closed? if yes, how was the device removed? what troubleshooting steps were attempted to open the device (e.G.Squeezing the closing trigger while simultaneously depressing the anvil release button, knife reverse switch, manual override)? did the jaws eventually open? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
During an unk procedure, the device cannot be opened anymore.No patient harm was reported.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 09/15/2025.D4: batch #247d84.Corrected data: h4.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one long60a device was returned with no apparent damage and with no reload present.The device was tested for functionality in the articulated position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.The event described could not be confirmed as the device performed as intended and it opened and closed without any difficulties noted.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number 247d84, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|