• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. 145-DEG PE 42MM HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. 145-DEG PE 42MM HUM LINER +2.5; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-42-03
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Joint Dislocation (2374)
Date of Event 12/20/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported in a shoulder clinical study approximately 7 months after a left shoulder replacement surgery a patient reported he dislocated his shoulder.X-rays were obtained and revealed disassociation of the polyethylene liner.As a result, the patient underwent a left shoulder revision of the humeral tray, humeral liner, glenosphere and glenosphere locking screw.The patient outcome was reported as resolved.No additional information is available.
 
Additional Manufacturer Narrative
D10: 320-15-03 - rs glen pla l post aug, 8 deg, left: (b)(6).320-10-05 - eque rev tray adap pl tray +5: (b)(6).320-06-42 - glenosphere 42mm: (b)(6).300-01-14 - equi hum stem primy pr fit 14mm: (b)(6).320-15-05 - eq rev locking screw: (b)(6).The reported event was unable to be confirmed due to limited information received from the customer.No device was returned for evaluation; further, photographs and/or radiograph images were not provided for review.Operative notes and/or medical records were not provided for review of usage/technique.A review of manufacturing data was unable to be performed as the lot information of the product involved in the event was not available.A definitive root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected/ additional information.The following sections were corrected/added: h6: component code, type of investigation.The revision reported was likely the result of dislocation and disassembly between the humeral liner and humeral tray.However, this cannot be confirmed and potential contributions from patient conditions, an unreported traumatic event, incomplete seating of the humeral liner during implantation, forceful contact between the liner and glenoid bone (impingement), or a combination of the above to the reported disassembly of the humeral liner cannot be determined as the devices were not returned for evaluation and images/radiographs were not provided.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
145-DEG PE 42MM HUM LINER +2.5
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct
gainesville, FL 32653
MDR Report Key22743227
Report Number1038671-2025-02703
Device Sequence Number8419340
Product Code KWT
UDI-Device Identifier10885862086709
UDI-Public10885862086709
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-42-03
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/26/2024
Supplement Date Received by Manufacturer01/13/2026
Initial Report FDA Received Date08/07/2025
Supplement Report FDA Received Date01/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age64 YR
Patient SexMale
Patient Weight93 KG
Patient RaceWhite
-
-