| Model Number |
LF1944 |
| Medical Device Problem Codes |
Energy Output Problem (1431); Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant product: vlft10gen; vlft10gen ft series energy platformx1; (serial no: (b)(6)).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the vein harvesting procedure, the device¿s handle became stuck in the pulled position and required significant effort and multiple attempts to return it to the open position.There was an inadequate or partial sealing, with evidence of energy delivery and end tones heard but no regrasp alert.Bleeding occurred after cutting, following an attempted seal of veins with tissue less than 7 mm thick.The jaws were cleaned several times during the procedure.Another device was used successfully with the same generator.
|
| |
|
Event or Problem Description
|
|
It was reported that during a vein harvesting procedure, the device's handle became stuck in the pulled position and required significant effort and multiple attempts to return it to the open position.There was an inadequate or partial sealing, with evidence of energy delivery and end tones heard but no regrasp alert.Bleeding occurred after cutting, following an attempted seal of veins with tissue less than 7 mm thick.The jaws were cleaned several times during the procedure.Another ligasure device was used successfully with the same generator.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: b5, d9, g3, h3, h6.H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that there was an inadequate or partial sealing and the jaws was difficult to open.The reported issues could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|