|
It was reported that, during use, the electrodes of a swan ganz pacing catheter got stuck in a 3rd party sheath.Patient with a right intrajugular temporary wire in place came to the electrophysiology lab for a device placement.After the pacemaker procedure was completed, temporary wire was pulled under fluoroscopy.After unlocking the valve of the sheath, the wire was being pulled while leaving the sheath in place.As the temporary wire was close to exiting the sheath, resistance was felt and prompted the staff to pull slightly harder.This caused the electrodes to break off.To prevent it from migrating into the patient's circulatory system, staff grabbed a 10cc syringe and pulled the blood back into the side port then immediately pulled the sheath to remove it from the patient.While pressure was placed on the intrajugular site, the circulating nurse began inspecting the sheath.The nurse cut the sheath open and was able to find the electrode portion of the wire adhered to the locking portion of the sheath.However, the nurse was not able to remove it as it was stuck.The physician also visually inspected it and confirmed that nothing was left inside the patient.
|