W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT321414 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Clinical Endoleak (4843)
|
| Date of Event |
10/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following information was reported to gore: on (b)(6) 2024, the patient underwent endovascular treatment for an abdominal aortic aneurysm using the gore® excluder® conformable aaa endoprosthesis, and the gore® excluder® aaa endoprosthesis.During deployment of the proximal portion of the trunk-ipsilateral leg endoprosthesis, movement to distal side occurred.Repositioning was performed using the constraining system.On completion angiography, findings suggestive of a proximal type i endoleak were observed.Balloon dilation was attempted, but no improvement was noted.Due to poor vascular condition at the aortic neck, no further treatment was performed.On an unknown date, during follow-up, a proximal type i endoleak or a type ii endoleak was suspected.On (b)(6) 2025, the patient underwent reintervention to treat the endoleaks.For the proximal type i endoleak, coil embolization was performed outside the proximal portion of the trunk-ipsilateral leg endoprosthesis.Embolization was also performed to treat type ii endoleaks originating from the lumbar artery and the left internal iliac artery.Additionally, to repair a right internal iliac artery aneurysm (outside the treatment area), embolization and additional stent graft placement were carried out.Physician's comment: during the initial procedure, no additional treatment was performed for the proximal type i endoleak due to poor vascular condition.However, placing an additional stent graft would have been preferable.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.
|
| |
|
Search Alerts/Recalls
|
|
|