| Catalog Number |
8065750852 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A non-health care professional reported that phacoemulsification tip was clogged at an unknown timing.Procedure details were unknown.There was no patient impact reported.Additional information received that event occurred during cataract surgery with no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
No technical inquiries were received from the customer.A sample was not received at the manufacturing site for evaluation for the report; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information received, the investigation was unable to identify the root cause or origin of the reported event.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time.All phacoemulsification tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|