• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING; SHOULDER PROSTHESIS, REVERSE CONFIGURATION Back to Search Results
Catalog Number 115370
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/22/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: concomitant medical products, part (lot): -(b)(4).Multiple mdr reports were filed for this event.Associated reports are available for the applicable concomitant products below in the d10 narrative.The customer has not indicated whether the product will be zimmer biomet for investigation, and the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported a patient had an initial right total shoulder arthroplasty followed by a revision on approximately 6 years post-implantation due to pain.Subsequently, reported an unknown implant broke and underwent a second revision approximately 7 years posterior.The bearing and humeral tray were revised and no additional information has been provided.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated/corrected: b4; b5; d2; d4; g1; g3; g6; h1; h2; h3; h6; h10.The reported event is not confirmed.Visual examination of the provided videos identified the patient lifting heavy weights after the noted revision.However, no product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were provided and reviewed by a health care professional regarding the revision in the separate linked complaint.Medical records were not provided for the revision accounted for in this complaint.The videos provided of the patient performing heavy weightlifting were from on two specific days.Unable to determine at this time if the patient performed heavy weightlifting prior to revision.Therefore, a definitive root cause cannot be determined.If any further information is found that would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMPREHENSIVE REVERSE SHOULDER HUMERAL TRAY WITH LOCKING RING
Common Device Name
SHOULDER PROSTHESIS, REVERSE CONFIGURATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key22748410
Report Number0001825034-2025-02398
Device Sequence Number18531244
Product Code PHX
UDI-Device Identifier00880304543201
UDI-Public(01)00880304543201(17)270503(10)271640
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K193373
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/18/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number115370
Device Lot Number271640
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2025
Supplement Date Received by Manufacturer11/18/2025
Initial Report FDA Received Date08/08/2025
Supplement Report FDA Received Date11/19/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/03/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age63 YR
Patient SexMale
Patient Weight78 KG
-
-