• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUREFIRE SCORPION NEEDLE; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number SUREFIRE SCORPION NEEDLE
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Foreign Body In Patient (2687); Insufficient Information (4580)
Date of Event 07/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
On 07/21/2025, it was reported by an arthrex subsidiary employee via (b)(4) that an ar-13991n surefire¿ scorpion¿ needle broke.This occurred during a case.There was no additional information provided, and additional information has been requested.
 
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Additional Manufacturer Narrative
Additional information: b5, g3, h6.
 
Event or Problem Description
Additional information was provided on (b)(6) 2025 where it was stated this event occurred during a shoulder arthroscopy when the doctor went to activate the forceps the device broke.A scorpions forceps was used with the needle.The patient had a normal bone quality that was prepped using a shaver blade.Another needle was opened to proceed with the case.A fragment remains in the patient.There was no delay in the procedure.
 
Event or Problem Description
Additional information was provided on 08/14/2025 where it was stated this event occurred during a shoulder arthroscopy when the doctor went to activate the forceps the device broke.A scorpions forceps was used with the needle.The patient had a normal bone quality that was prepped using a shaver blade.Another needle was opened to proceed with the case.A fragment remains in the patient.There was no delay in the procedure.
 
Additional Manufacturer Narrative
Additional info: b5, g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude the most likely cause.The reported failure is caused by a user error, specifically applying excessive force to pass the needle through fibrous/calcific tissue and/or ¿dry,¿ repetitive actuations outside the surgical site, as stated in the directions for use (dfu-0392-sub-eo warning for scorpion products).To help avoid needle breakage and potential patient injury: do not resterilize or reuse the scorpion suture passer needle.Avoid striking bone or using excessive force to pass the needle through fibrous/calcific tissue.If excessive force is encountered, replace the needle, reposition the device and pass in a different area.Ensure a secure grasp of tissue to prevent the needle from buckling under the tissue.Avoid "dry" repetitive actuations outside the surgical site.Abrasion of the needle surface can occur that may inhibit proper function.The complaint allegation could not be confirmed.The device was not received for evaluation, and no supporting evidence of the reported failure was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUREFIRE SCORPION NEEDLE
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22748801
Report Number1220246-2025-03345
Device Sequence Number15871197
Product Code MDM
UDI-Device Identifier00888867015920
UDI-Public(01)00888867015920(11)241014(17)290930(10)15331441
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/28/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUREFIRE SCORPION NEEDLE
Device Catalogue NumberAR-13991N
Device Lot Number15331441
Was Device Available for Evaluation? No
Distributor Facility Aware Date07/21/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by Manufacturer08/14/2025
08/14/2025
Initial Report FDA Received Date08/08/2025
Supplement Report FDA Received Date08/27/2025
08/28/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-