On 07/22/2025, a sales representative reported via phone that an ar-18716-13 cortical screw broke.This occurred during a clavicle orif procedure on (b)(6) 2025.The fragments were retrieved.The patient had a normal bone quality.There was no delay in the procedure, and no additional anesthesia was administered.The case was completed using a non-arthrex device.
|
|
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional field information, arthrex was able to determine the most likely cause.The most likely cause(s) for the reported failure can be attributed to user error, improper bone preparation, misaligned insertion, or excessive forces through leveraging/prying the device.The complaint allegation could not be confirmed, as the device was not received for evaluation, and no evidence of the reported failure was provided.
|