Additional information: d9, g3, h3, h6.-one unpackaged ar-2652cr was received for investigation.-upon visual inspection, the plate was broken off at one of the two oblong holes located in the middle of the plate.The plate was bent, nicks were observed around the edges, and scratches on it.-dimensional inspection was conducted in accordance with drawing (b)(4), and all measured values were found to be within specified tolerances.(refer to dimension results for details.) -functional testing was not performed due to the damage to the device.Per surgical technique, clavicle plate and screw system plate selection select the appropriate clavicle plate to match the patient¿s anatomy.Aluminum clavicle plate sizing templates may be used to help determine the appropriate implant.The plates are precontoured to reduce the need to bend.If contouring the plate is necessary, use the appropriate plate benders.Do not bend the plate near the locking holes as this may distort the threads and inhibit proper screw insertion.Repeated bending of the plate at the same location or by creating excessive acute angles may potentially lead to plate fatigue, failure, and/or breakage in situ.Dfu-0192-eo-e.Warnings 7.Preoperative and operating procedures, including knowledge of surgical techniques and proper selection and placement of the device, are important considerations in the successful utilization of this device.The appropriate arthrex delivery system is required for proper implantation of the device.The most likely cause for the reported failure can be attributed to user error of the device due to improper surgical technique and applying excessive mechanical during insertion.-refer to investigation photos.-complaint allegation is confirmed.
|