• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. 3.5MM PEEK SELF-PUNCHING PUSHLOCK; NON-DEGRADABLE FIXATION FASTENER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. 3.5MM PEEK SELF-PUNCHING PUSHLOCK; NON-DEGRADABLE FIXATION FASTENER Back to Search Results
Model Number 3.5MM PEEK SELF-PUNCHING PUSHLOCK
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/22/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation is in process.A follow-up report will be provided upon availability of additional information.
 
Event or Problem Description
It was reported that during a trochleaplasty surgery the device broke.All broken parts were retrieved from the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6.The complaint device was not received for evaluation.Based on the information provided, which may include the device (if available and returned), pictures, videos, event descriptions, and any additional field information, arthrex was able to determine the most likely cause.The most likely cause of the reported failure is user error, including improper bone preparation and incorrect surgical technique during device application.These factors may have contributed to the malfunction or compromised the integrity of the implant during the procedure.The complaint allegation could not be confirmed, as the device was not received for evaluation, and no evidence of the reported failure was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.5MM PEEK SELF-PUNCHING PUSHLOCK
Common Device Name
NON-DEGRADABLE FIXATION FASTENER
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key22751518
Report Number1220246-2025-03378
Device Sequence Number9558091
Product Code MBI
UDI-Device Identifier00888867395268
UDI-Public(01)00888867395268(11)230717(17)280630(10)15117034
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K251145
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 11/21/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number3.5MM PEEK SELF-PUNCHING PUSHLOCK
Device Catalogue NumberAR-1926PSSP-1
Device Lot Number15117034
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/22/2025
Supplement Date Received by Manufacturer11/14/2025
Initial Report FDA Received Date08/08/2025
Supplement Report FDA Received Date11/21/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/17/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-