MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
SHPX650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
09/06/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with unspecified mentor gel breast implants experienced rupture on the right side post-operatively, which was confirmed during a physical exam.As a result, the patient underwent device removal.
|
| |
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.Since no lot number was provided, no manufacturing record evaluation could be performed.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.D4: udi: as the catalog/model number was not provided, the (01)gtin is not available.Reason for device explant and/or reoperation: rupture.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On aug 7, 2025, additional information was received indicating the patient's race is caucasian.The patient's age at the time of event was 40 years old.The date of implantation was identified as jul 26, 2024.The date of event was identified as (b)(6) 2024.The date of explantation was identified as (b)(6) 2025.The impacted product was identified as a mentor memorygel xtra breast implant, catalog number shpx650, serial number (b)(6).A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A photo of the device was returned for evaluation.The evaluation was completed on sep 18, 2025.Photo evaluation summary: upon visual evaluation of the image provided in the complaint, the implant was observed ruptured.As the product involved in this complaint was discarded, the device could not be analyzed according to our procedures.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On nov 10, 2025, it was noticed d6b.Explantation day was incorrect in a previous report.The field has been updated appropriately.Manufacturer¿s reference number: (b)(4).
|
| |
|