• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN AVITUM ITALY S.P.A. PINNACLE®; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN AVITUM ITALY S.P.A. PINNACLE®; CONTAINER, I.V. Back to Search Results
Catalog Number 2112350
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow-up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the b.Braun global affiliate: event 3 when did the failure occur: before application reason of complaint: presencia de contaminante (cabello) presence of contaminant (hair).
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number(b)(4).Event 3.Five (5) samples in original closed packaging and a photo were submitted to the manufacturer for evaluation.Through visual examination of the photo, a thin dark filament comparable to human hair was observed.Through visual examination of the samples, the presence of a human hair on the first bag was observed.The reported defect was observed; the complaint is considered confirmed.It was noted that the contamination is outside the product external to the fluid path.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.The defect is a result of a failure during the packaging phase.The production site meets clean room standard; hygiene regulations and rules of dressing are in place.The production department will be informed of this complaint and defect.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINNACLE®
Common Device Name
CONTAINER, I.V.
Manufacturer (Section D)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT  41037
Manufacturer (Section G)
B. BRAUN AVITUM ITALY S.P.A.
via xxv luglio 11
mirandola, 41037
IT   41037
Manufacturer Contact
jonathan severino
901 marcon blvd
allentown, PA 18109
4847197287
MDR Report Key22751617
Report Number9681240-2025-00057
Device Sequence Number8607208
Product Code KPE
UDI-Device Identifier04046964200793
UDI-Public(01)04046964200793(11)241201(17)271130(10)24M02
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K041415
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/02/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2112350
Device Lot Number24M02
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/21/2025
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/08/2025
Supplement Report FDA Received Date09/02/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/01/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-