| Catalog Number |
TSV4H8 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Unspecified Infection (1930); Inflammation (1932); Osteolysis (2377)
|
| Date of Event |
03/13/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A summary investigation has been completed for peri-implantitis events recognizing that a definitive root cause cannot be identified due to a wide range of external factors (non-design or manufacturing related), including medical conditions (e.G., diabetes, poor bone quality, etc.) / patient habits (e.G., smoking) and surgical technique.Previously completed investigations for these events have not identified any signals indicating potential non-conformances affecting the manufacturing and sterilization processes.Furthermore, the probability of a manufacturing or design defect that might lead to peri-implantitis occurring and escaping the available detections has been assessed and found remote and almost nonexistent.Should additional information be received which indicates that the device may have caused or contributed to the event, an additional report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the implant at tooth site #27 was removed due to peri-implantitis.The initial diagnosis was periapical periodontitis (pap) in 27 dental implant.The extraction of this implant was scheduled for 3-6 months later, with patient's being informed about the treatment plan, costs, and prognosis.After obtaining informed consent and signing the medical authorization form, routine disinfection of the surgical area was performed.Under local infiltration anesthesia using 1.7ml of adetacaine adrenaline injection, the gums were incised to extract the 27 dental implant.The extracted tissues were carefully scraped, sutured, and compressed with pressure for 30 minutes.Symptoms as a result of the event: inflammation.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).Dhr review was completed for the subject lot number (b)(6).No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Sterilization record was reviewed and verified to have passed all sterilization activities with no issues or nonconformities identified.Complaint history review was performed for the reported lot number 1274536 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental : medical : peri-implantitis¿.
|
| |
|