• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE Back to Search Results
Model Number BF-UC180F
Medical Device Problem Code Poor Quality Image (1408)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/16/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned to olympus for inspection when reportable malfunction was observed.Based on the results of the investigation, the issue is attributed to stress/breakage of the ultrasound transducers, however, a definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was observed during olympus' device inspection that the bronchofibervideoscope exhibited degradation of the 3d image.There were no reports of patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Common Device Name
ULTRASONIC BRONCHOFIBERVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key22756412
Report Number3002808148-2025-12518
Device Sequence Number8607052
Product Code PSV
UDI-Device Identifier04953170356353
UDI-Public04953170356353
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K070983
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/08/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberBF-UC180F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/16/2025
Initial Report FDA Received Date08/08/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-