MENTOR TEXAS MENTOR MEMORYGEL BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
|
Back to Search Results |
|
| Catalog Number |
3505800BC |
| Medical Device Problem Codes |
Material Rupture (1546); Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Code |
Deformity/ Disfigurement (2360)
|
| Date of Event |
06/14/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: breast prosthesis rupture.H6 medical device problem code a27: bubbles in implant.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a breast reconstruction revision procedure with mentor memorygel breast implant 800cc gel breast prostheses that both ruptured post implantation.Mri results indicated right breast prosthesis rupture and showed bubbles in the left breast prosthesis.The doctor wanted to get the mri results reviewed for ¿micro rupture¿ in the left breast prosthesis.As a result, the patient underwent bilateral explantation and replacement with mentor memorygel breast implant 800cc gel breast prostheses on (b)(6) 2025.This medwatch report is for the patient¿s left breast prosthesis.
|
| |
|
Additional Manufacturer Narrative
|
|
On 12-aug-2025, the mentor failure analysis lab received the device for evaluation.On 14-aug-2025, mentor completed the investigation on the suspect medical device.Device evaluation summary: according to the information received, the breast prosthesis had bubbles in it and the surgeon wanted the mri results review for ¿micro rupture¿.The product was returned to mentor for evaluation.Mentor conducted visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned device revealed that bubbles on the breast implant.Leak testing was performed, in accordance with mentor procedures, and it identified a tear on the anterior view, measuring approximately less than 0.1 cm; air could have passed through the tear and originated the bubbles reported.In addition, no other areas of gel exposure were detected during the analysis.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|