• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARGON MEDICAL DEVICES TLAB TRANSJUGULAR LIVER BIOPSY SYSTEM; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARGON MEDICAL DEVICES TLAB TRANSJUGULAR LIVER BIOPSY SYSTEM; INTRODUCER, CATHETER Back to Search Results
Model Number TL-19
Medical Device Problem Code Defective Component (2292)
Health Effect - Clinical Code Skin Tears (2516)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The sample is indicated as available for return.As of the date of this report, the sample has not been returned.A follow-up report will be provided once the device has been received and reviewed.
 
Event or Problem Description
Reporter states "i do have a defective tlabs kit where the plastic sheath covered over the top of the needle and consequently the md was cut trying to unsheathe the needle tip." additional information has been requested.
 
Additional Manufacturer Narrative
We conducted a thorough review of the manufacturing and inspection records for this lot and found no deviations or non-conformances.Unfortunately, we did not receive any samples for evaluation, nor did we receive any photographic or visual evidence that would have allowed us to review the allegation in detail.Without this necessary evidence, we are unable to perform a thorough investigation or determine a root cause and corrective action at this time.As a result, no corrective action is required at this point.However, if samples are returned at a later date, we would be happy to reopen this complaint for re-evaluation.Additional information provided in h.3., h.6.And h.11.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TLAB TRANSJUGULAR LIVER BIOPSY SYSTEM
Common Device Name
INTRODUCER, CATHETER
Manufacturer (Section D)
ARGON MEDICAL DEVICES
1445 flat creek road
athens TX 75751
Manufacturer Contact
kelly mayer
1445 flat creek road
athens, TX 75751
9036759321
MDR Report Key22767517
Report Number0001625425-2025-01062
Device Sequence Number8134263
Product Code DYB
UDI-Device Identifier00886333218516
UDI-Public(01)00886333218516(17)280328(10)3707432
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K022634
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/30/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTL-19
Device Lot Number3707432
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/10/2025
Supplement Date Received by Manufacturer10/01/2025
Initial Report FDA Received Date08/09/2025
Supplement Report FDA Received Date10/30/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-